Testing Soil Sensors Against a Reference

2 min read
ValidationMethodology

Until a reading has been compared against something trustworthy, it is a number the device produced, not a measurement of the soil. This entry covers the first structured comparison we ran.

Setup

We selected twelve points across two fields with visibly different drainage characteristics. At each point we took a probe reading, then extracted a sample from the same location and depth within a few minutes, and sent those samples for accredited laboratory analysis.

The comparison covers pH and electrical conductivity. Moisture was logged alongside but validated separately, against oven-dry gravimetric measurement.

Twelve points is a small sample. It is enough to detect a large systematic problem and not enough to characterise accuracy, which is the correct way to read what follows.

Results, stated carefully

pH tracked the laboratory values reasonably well. Direction and relative ordering were consistent — points the laboratory ranked more acidic, the probe also ranked more acidic. Absolute agreement was weaker, with a consistent offset that varied by field.

That offset pattern is informative. A consistent offset suggests calibration, which is correctable. Scattered disagreement would have suggested something worse.

EC disagreed substantially, and moisture explained most of it. Once we grouped readings by measured moisture content, the relationship between probe EC and laboratory EC tightened considerably within each group. Across groups it did not exist at all.

This is not a discovery — it is documented soil physics. What we learned was the magnitude in our specific conditions, which is what determines whether a correction is a minor adjustment or the central problem.

The finding that changed our priorities

We ran three readings at each point rather than one. The spread between repeats at the same point was, for several points, larger than the gap between our mean reading and the laboratory value.

That is worth restating plainly: our sampling variability was larger than our sensor's disagreement with the reference. Improving the electronics would not have improved the result. Improving how the probe is placed in the soil would.

This is why the version two changes are mechanical.

What this does not tell us

  • Nothing about accuracy in soils unlike these two fields.
  • Nothing about stability over a season — this was a single day.
  • Nothing about nutrient measurement, which was not part of the comparison.
  • Nothing about whether the probe survives being left in a field.

We would need considerably more data before quoting an accuracy figure, and we are not going to quote one until we have it.