Our original assumption was that a probe could be calibrated once, at assembly, and then trusted in the field. That assumption did not survive contact with a second soil type.
The problem
Capacitive moisture sensing responds to the dielectric permittivity of the surrounding medium. Soil is not one medium — it is mineral particles, water, air, and organic matter in a ratio that varies with texture. A calibration curve fitted on a sandy loam systematically misreads a clay soil, because the same water content produces a different dielectric response.
We measured this. Applying our original curve to a heavier soil produced moisture readings that were consistently wrong in the same direction, by enough to matter for any EC correction that depended on them.
Since EC correction depends on moisture, and our EC interpretation depends on that correction, an error here propagates into everything downstream. That made it worth solving before anything else.
What we changed
Calibration is now a per-site procedure rather than a factory step.
The routine takes two reference points at a location: one at or near saturation, one at air-dry, using soil from that site. The device fits a curve between them and stores it against the site identifier. Every subsequent reading at that site is interpreted through its own curve.
The stored curve carries a timestamp. Readings taken more than a set interval after the last calibration are flagged in the record rather than silently treated as current — a stale calibration should be visible in the data, not discovered later.
What this costs
Honesty about the trade-off: this makes the device meaningfully harder to use. A per-site calibration is a real task requiring soil samples and some care, where previously there was nothing to do.
We think it is the right trade. An easy device that produces unreliable numbers does not save anyone time; it costs them a season. But it does move work onto the user, and reducing that burden — ideally by inferring texture rather than requiring the reference points — is now on the list.
Still unresolved
- Temperature compensation is applied but has not been validated across a wide enough range.
- We do not yet know how quickly a site calibration goes stale in practice.
- The two-point fit assumes a curve shape that may not hold for soils with high organic matter. Untested.